Procedure
-
Turn on the spectrophotometer GENESYS20
- Allow the instrument to warm up for at least 10 minutes if required by the manufacturer.
-
Set wavelength
- Adjust the spectrophotometer to 640 nm.
-
Prepare the blank
-
Fill a clean cuvette with 1 mL (or the required volume) of the same sterile medium used to grow the culture.
-
Wipe the outside of the cuvette with a lint-free tissue to remove fingerprints or droplets.
-
-
Blank the instrumen
-
Insert the blank cuvette into the holder, aligning the transparent sides with the light path.
-
Close the lid and press “Blank” or “Zero”.
-
Wait until the display reads 0.000 absorbance (or 100% transmittance).
-
-
Measure the sample
-
Mix the microbial culture gently by inversion or pipetting (do not vortex to avoid cell lysis or bubbles).
-
Transfer 1 mL of the culture into a clean cuvette.
-
Wipe the cuvette and place it in the same orientation as the blank.
-
Record the OD₆₀₀ value.
-
-
Dilute if necessary
-
If the OD₆₀₀ value exceeds 0.5, dilute the sample (e.g., 1:10 in fresh medium), mix, and remeasure.
-
Multiply the measured OD₆₀₀ by the dilution factor to obtain the true OD₆₀₀.
-
-
Clean up
-
Rinse cuvettes thoroughly with distilled water and let them air-dry or store according to lab policy.
-
Turn off the spectrophotometer if not in continuous use.
-
Notes
-
Always use the same type of cuvette (plastic or glass) for blank and samples.
-
Avoid bubbles and residual droplets—they can alter the reading.
-
Record all values in a lab notebook or data sheet, including dilution factors and time points.
1. The origin of OD₆₀₀
-
OD₆₀₀ (600 nm) became the standard wavelength for measuring bacterial growth, particularly E. coli, because:
-
Early photoelectric colorimeters and spectrophotometers (mid-20th century, e.g., Klett-Summerson colorimeter) used filters centered around 600 nm (“Klett filter #66” was 600 ± 20 nm).
-
At 600 nm, visible light scattering by bacterial cells is high, but absorption by medium components (especially proteins and riboflavin) is minimal.
-
It provided a convenient linear correlation between cell density and light scattering up to OD ≈ 0.8–1.0.
-
Consequently, OD₆₀₀ was widely adopted in bacterial growth protocols, and it became embedded in textbooks and instrument presets.
-
2. Why OD₆₄₀ (or OD₆₃₀–₆₄₀) is sometimes used
-
Yeast (e.g., Saccharomyces cerevisiae) cells are larger and more refractile than bacteria; they scatter light differently.
-
Early yeast researchers found that measurements at slightly longer wavelengths (620–640 nm):
-
Reduced multiple scattering (less overestimation at high cell densities).
-
Improved linearity between OD and actual cell dry weight or concentration.
-
Minimized interference from colored media or metabolic pigments (like flavins or residual YPD coloration).
-
-
Some instruments (especially older spectrophotometers in yeast labs) came equipped with a 640 nm or 620 nm filter, not 600 nm, simply because they were originally designed for turbidity or colorimetric assays (e.g., McFarland standards, enzymatic reactions).
-
As a result, OD₆₄₀ became the norm in many yeast laboratories — particularly in Europe (e.g., Scandinavian and German microbiology traditions).
3. Practical differences
-
Both wavelengths measure scattered light, not true absorption.
-
The actual difference in measured OD between 600 and 640 nm is typically small (≈ 5–10%), but not negligible if one is comparing across studies.
-
Empirically, some labs use correction factors: [ \text{OD}{600} ≈ 1.1 × \text{OD}{640} ] (though this varies depending on the spectrophotometer and culture conditions).
4. In the literature
-
E. coli and most bacteria → OD₆₀₀ (historical Klett tradition, standard in molecular biology).
-
S. cerevisiae and other yeasts → OD₆₄₀ or OD₆₃₀ (to improve linearity with biomass and because of larger cell size/scattering behavior).
-
Microalgae and colored cells → even longer wavelengths (e.g., 680 nm) to minimize chlorophyll absorption overlap.
5. Bottom line
The difference is not biochemical, but optical and historical:
OD₆₀₀ comes from bacterial tradition (Klett filter #66). OD₆₄₀ emerged from yeast work to improve scattering linearity and reduce medium interference.
Both are valid, as long as:
-
You consistently use one wavelength within your lab.
-
You report it explicitly (e.g., “Cell density was monitored at OD₆₄₀”) when publishing.